EZTestŪ is a self-contained biological indicator for monitoring steam / flash steam, ethylene oxide, hydrogen peroxide, or propylene oxide sterilisation. EZTestŪ contains bacterial spores on a filter-paper carrier packaged within a small, thermoplastic culture tube. Inside the thermoplastic culture tube is a sealed-glass ampoule of specially-formulated soybean casein digest culture medium containing a colour indicator which turns a dramatic yellow when spores grow. EZTestŪ is easy-to-use and produces visual results within 24 hours (steam, hydrogen peroxide, or propylene oxide) or 48 hours (ethylene oxide) without laboratory transfers.
|
||
![]() |
![]() |
![]() |
![]() |
||
![]() |
![]() |
![]() |
![]() |
|
![]() |
![]() |
|
||||||||
![]() |
![]() |
![]() |
Copyright Ati Atlas Ltd 2009 |
||
![]() |
![]() |
![]() |
EZTestŪ Steam/Flash |
||||
Spore |
Incubation |
Population |
Unit |
Product Code |
|
24 Hours at 55-60°C |
10^5 |
100 |
EZS/5 |
10^5 |
25 |
EZS/525 |
||
10^6 |
100 |
EZS/6 |
||
10^6 |
25 |
EZS/625 |
||
![]() |
EZTestŪ Gas (Ethylene Oxide) |
||||
Spore |
Incubation |
Population |
Unit |
Product Code |
B. atrophaeus |
35-39°C |
|
100 |
EZG/6 |
25 |
EZG/625 |
|||
![]() |
EZTestŪ H202 (HYDROGEN PEROXIDE) |
||||
Spore |
Incubation |
Population |
Unit |
Product Code |
G. stearothermophilus |
55-60°C |
10^5 |
|
EZH/5 |
10^6 |
EZH/6 |
|||
![]() |
EZTestŪ PPO (PROPYLENE OXIDE GAS) |
||||
Spore |
Incubation |
Population |
Unit |
Product Code |
|
55-60°C |
|
|
|
![]() |
EZTestŪ Availability
Storage : Room temperature.
Expiration: 24 months from date of manufacture.
Certification: EZTestŪ biological indicators are traceable to a recognised culture collection and certified for population, D-value, Z-value and performance.